Application Note #60 – Turbidity
Application Note #60 – Turbidity
Updated September 2017
Introduction:
Turbidity is defined as the amount of suspended particles in a solution, measured in nephelometric turbidity units, (NTU). Turbidity is used as a general indicator of the quality of water, along with colour and odour. A large concentration of particles in drinking water is undesirable as it causes an unpleasant taste and appearance as well as reducing the ability of chemicals and radiation to neutralize disease causing bacteria and viruses.
Turbidity can be measured by the disk method, by the Jackson candle method, or more accurately and precisely by the nephelometric method. The disk method is appropriate for rough turbidity measurements of lakes and rivers and involves lowering a coloured disk into the water and recording the distance at which the disk disappears from view. The Jackson candle method uses a tube of glass with a candle underneath and recording the length of the tube that is filled when the candle disappears from view.
The most recent method, which is called the nephelometric method, involves shining light into a sample and recording the light scattered at 90 degrees to the light path. By also recording the light transmitted straight through the sample, the ratio of light scattered to that transmitted may be used to compensate for the interference seen in samples which have colour above 15 CU. There has been some debate over the wavelengths to be used for the nephelometric work. The US EPA approved source is a tungsten white light source while the international standard is an 800nm infrared light source. The turbidity meter described in this method uses a tungsten white light source to meet US EPA requirements. The alternative method is identical to the white light method, except that the meter uses an infrared light source that allows for more accuracy with colored water samples.
There are several standards requiring the measurement of turbidity levels in water. The US EPA has a maximum contaminant level (MCL) of 5 NTU for drinking and wastewaters, while the Canadian national standard for drinking water is an aesthetic objective (AO) of 5 NTU and a maximum acceptable concentration (MAC) of 1 NTU. As this method automatically, accurately and precisely measures turbidity down to 0.1 NTU it is well suited for monitoring the turbidity in water samples.
Conforms to: EPA 180.1, ASTM D1889, and Standard Method 2130 B (White Light); ISO 7027 and EN27027 (Infrared)
Sample: This method was developed for drinking, waste, and environmental water samples.