How can the Permanganate Index test be automated?

MANTECH is able to automate all aspects of the permanganate index test including automated sample heating, acid and reagent addition, and titration with calculation of results. The analysis can also be optimized for maximum throughput capacity through a batch setup where multiple samples at once are heated, prepared, and analyzed all within the timing specified in the ISO 8467 method. Overall, the benefit of the batching process is much quicker time per sample overall since you are making use of the 10 minute wait times between adding the reagents instead of the system just sitting and waiting.

 

The time per sample for single analysis is roughly 25 minutes per sample taking into account all the waiting times and about 3 minutes for the actual titration.

 

For the batch analysis, you get roughly 12 samples per batch and the batch takes roughly 60 minutes to complete, making it only about 5 minutes per sample.

 

Below are the steps both for single analysis and batch analysis:

 

Single Analysis:

  1. A blank analysis performing the steps 3-11 listed below, using DI water as the sample, is run and the blank endpoint volume is used in the calculation of sample results for the run.
  2. After titrating the blank sample, 5mL of 0.005M Sodium Oxalate is added to the titrated sample, then after a brief mixing period the titration is performed again. The endpoint volume is recorded as the standardization volume and is also used in the calculation of sample results for the run.
  3. 25mL of the sample volume is pipetted into each sample beaker, and sample protect caps are placed on the beakers.
  4. The Heating Element are turned on. Heating is to be applied keeping the sample temperature between 96-98°C during the following steps until the element is turned off.
  5. The Sample Protect cap is removed and 5mL of 2.0M Sulfuric Acid is pipetted into a beaker. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed on the beaker.
  6. 10-minute wait time.
  7. The Sample Protect cap is removed and 5mL of 0.002M Potassium Permanganate solution is added. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed back on the beaker.
  8. 10-minute wait time.
  9. The Sample Protect cap is removed and 5mL of 0.005M Sodium Oxalate solution is added. The stirrer turns on to mix, at which point the solution will briefly change color but eventually become colorless.
  10. 1-minute wait time.
  11. The solution is titrated with 0.002M Potassium Permanganate to a faint pink color. This color is measured with the photometric probe.
  12. Steps 5 to 11 are repeated for all remaining samples.

 

Batch Analysis:

  1. A blank analysis performing the steps 3-13 listed below, using DI water as the sample, is run and the blank endpoint volume is used in the calculation of sample results for the run.
  2. After titrating the blank sample, 5mL of 0.005M Sodium Oxalate is added to the titrated sample, then after a brief mixing period the titration is performed again. The endpoint volume is recorded as the standardization volume and is also used in the calculation of sample results for the run.
  3. 25mL of the sample volume is pipetted into each sample beaker, and sample protect caps are placed on the beakers.
  4. The Heating Element are turned on. Heating is to be applied keeping the sample temperature between 96-98°C during the following steps until the element is turned off.
  5. The Sample Protect cap is removed and 5mL of 2.0M Sulfuric Acid is pipetted into a beaker. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed on the beaker.
  6. Step 5 is repeated for multiple sample beakers in a row, until 10 minutes has elapsed since the acid was added to the first beaker. The set of beakers that get acid added to them within the 10 minute time become the first “batch”.
  7. After 10 minutes has elapsed, the system moved back to the first sample beaker in the batch.
  8. The Sample Protect cap is removed and 5mL of 0.002M Potassium Permanganate solution is added. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed back on the beaker.
  9. Step 8 is repeated for the remaining beakers in the batch determined from Step 6. Roughly 10 minutes should elapse from when the permanganate was added to the first beaker.
  10. After 10 minutes has elapsed, the system moved back to the first sample beaker in the batch.
  11. The Sample Protect cap is removed and 5mL of 0.005M Sodium Oxalate solution is added. The stirrer turns on to mix, at which point the solution will briefly change color but eventually become colorless.
  12. 1-minute wait time.
  13. The solution is titrated with 0.002M Potassium Permanganate to a faint pink color. This color is measured with the photometric probe.
  14. Steps 11-13 are repeated for the remaining samples in the first batch.
  15. After the first batch is complete, the system repeats steps 5-14 for all remaining samples in the run, creating as many “batches” as required to complete the full set of samples.