How does MANTECH account for temperature compensation and correction in pH measurements?

As the temperature of a solution changes, the actual pH changes. This is not an error of the probe or meter being used, but is the actual pH of the solution at that particular temperature. The temperature effect on the pH value is 0.003 pH units per oC away from 25oC, per pH units away from pH 7. This effect can be either negative or positive, depending on if the temperature is above or below 25oC, and if the pH is above or below pH 7. At 25oC and pH 7, there is no change in the pH value.

The chart below shows how the actual pH changes with temperature and pH, which allows you to correct the pH reading to 25oC. As an example, if a sample measured pH 5 at a temperature of 5oC, the chart indicates this a negative effect, therefore the sample pH would be 5 – 0.12 = 4.88pH corrected to 25oC:

temperature-effect-on-pH-chart

MANTECH software accounts for temperature by recording the temperature during calibration via a thermistor probe, then it corrects the pH reading at the time of measurement to account for the difference in measured temperature of the sample vs. the calibration buffers. The reported pH value is the corrected value @ the sample temperature. The equation below shows how the PC-Titrate software calculates pH:

 

pH-equation

EMeas  =  Voltage measure by electrode at the time of titration (mV)

EInt  = Voltage of the Intercept value calculated from the calibration equation. (mV)

Slope = slope of the line calculated from the calibration equation. (mV)

TMeas = temperature measured at the time of pH measurement (K degrees)

TCal = temperature taken at the time of last calibration. (K degrees)

 

MANTECH also offers the option to report pH values corrected to a set temperature, such as 25°C. This is implemented in the software as an additional step after the sample pH is recorded. MANTECH offers this as a feature for new MT-Series systems, or as an upgrade to existing MANTECH systems.

How do MANTECH systems measure and record temperature?

The standard method for measuring temperature is with a USB-connected stainless steel thermistor probe. The specifications for this probe are below:

  • Communication: USB controlled by MANTECH Pro Software
  • Sheath Material: Super OMEGACLAD XL/ Stainless Steel/Inconel
  • Temp Range: 0 to 1038 Degree C.
  • Diameter: 0.062in
  • Length: 6 inch
  • Measurement to 3 decimal places
  • Autocalibration feature in MANTECH Pro software

Alternatively, when measuring conductivity, one can measure the temperature directly from the conductivity probe. This temperature is also displayed directly on the conductivity meter screen.

MANTECH has the ability to use either of these measurement methods for pH temperature compensation, depending on customer preference and system configuration.

MANTECH temperature sensors are available in both PT1000 and 10kNTC styles. PT1000 sensors have a linear positive-slope relationship between resistance and temperature with a resistance of 1000 ohms at 0°C, and 10kNTC sensors have a curved negative-slope relationship between resistance and temperature with a resistance of 10,000 ohms at 25°C. Both these models of temperature sensor perform well and comparably within the 0-100°C temperature range that liquid samples exist in. If you require a specific type of temperature sensor, please feel free to let MANTECH know and we will accommodate.