Frequently Asked Questions

MT Analyzer

General Information

How does MANTECH perform automated in-line filtering of samples?

The MANTECH color system can be set up with a multiple in line filters in parallel between the autosampler and the spectrophotometer. The system pumps sample through the filter and automatically switches from one filter to the next at the specified times. This can either be determined by a set number of samples per filter, or by a flow detection system where there is a sensor detecting when the flow through the filter drops below a certain threshold, at which point the system automatically switches to the next fresh filter and continues the run.

A general example of the run sequence using flow detection would be:

  • Sample 1 runs through filter 1, flow does not decrease below the threshold, no change of filter
  • Sample 2 runs through filter 1, flow does not decrease below the threshold, no change of filter
  • Sample 3 runs through filter 1, flow decreases below threshold for clogged valve, system switches to filter 2
  • Sample 4 runs through filter 2, flow does not decrease below the threshold, no change of filter
  • Sample 5 runs through filter 2, flow decreases below threshold for clogged valve, system switches to filter 3
  • Sample 6 runs through filter 3, flow does not decrease below the threshold, no change of filter
  • Etc. for remainder of samples, or until all filters are clogged up
    • If all filters become clogged up, run pauses and informs user to switch the filters and start back up

Get Smarter Solutions, Today.

Our innovative water and wastewater analyzers have faster, safer processes, with fewer hazardous reagents. Start using MANTECH solutions for more accurate data that drives responsible outcomes.

Contact Us