Frequently Asked Questions

MT Analyzer

General Information

How do MANTECH systems measure and record temperature?

The standard method for measuring temperature is with a USB-connected stainless steel thermistor probe. The specifications for this probe are below:

  • Communication: USB controlled by MANTECH Pro Software
  • Sheath Material: Super OMEGACLAD XL/ Stainless Steel/Inconel
  • Temp Range: 0 to 1038 Degree C.
  • Diameter: 0.062in
  • Length: 6 inch
  • Measurement to 3 decimal places
  • Autocalibration feature in MANTECH Pro software

Alternatively, when measuring conductivity, one can measure the temperature directly from the conductivity probe. This temperature is also displayed directly on the conductivity meter screen.

MANTECH has the ability to use either of these measurement methods for pH temperature compensation, depending on customer preference and system configuration.

MANTECH temperature sensors are available in both PT1000 and 10kNTC styles. PT1000 sensors have a linear positive-slope relationship between resistance and temperature with a resistance of 1000 ohms at 0°C, and 10kNTC sensors have a curved negative-slope relationship between resistance and temperature with a resistance of 10,000 ohms at 25°C. Both these models of temperature sensor perform well and comparably within the 0-100°C temperature range that liquid samples exist in. If you require a specific type of temperature sensor, please feel free to let MANTECH know and we will accommodate.

Other

Support and Troubleshooting

Hardware

How do I clean my conductivity probe?

For water soluble contaminants, rinse probe in deionized (DI) water.  If ineffective, soak probe in warm DI water with household detergent for 15 – 30 minutes. For oil-based contaminants, rinse probe in ethanol or acetone for short (5-minute) periods.After cleaning, rinse probe in DI water to remove residual cleaning reagents. Perform a meter calibration before proceeding with sample analysis.

How to retrieve cell constant from consort conductivity meter?

The cell constant can be retrieved from the GLP report, directly found on the meter. Press CAL, then press the Down Arrow key to select GLP, and press OK. Press OK again on Show Report.This will bring up the calibration information was the performed on the meter, including the cell constant value. You need to scroll down, using the Down Arrow key, to find the section called “Calibration”, where CC = Cell Constant (cm-1). Note: Scrolling to the bottom will show the previous cell constant.

MANTECH Pro Software™

How to perform a meter calibration for conductivity?

MANTECH Pro Software™ automatically performs a single-point calibration without requiring the USER to interact with the meter. A calibration method will not be required. Before running your calibration script, ensure that the meter settings is configured with your probe.

The following steps will outline how to calibrate the meter:

For Automated systems

  1. Add in your tray file, rinsing, move to tube location and jogging movement steps, prior to the calibration step.
  2. Add a ‘Timing’ action of 30 seconds. This will allow the meter to stabilize the conductivity reading.
  3. Add the ‘Conductivity Meter Calibrate’ action and select the conductivity meter configured in Hardware Setup.
  4. Complete the script, configure the Run Configuration table, and click ‘Go to Run’ to start the meter calibration.

For Manual systems

  1. Select ‘Misc.’ and type in instructions for analyst to manually dip the probe into the standard solution 3 times.
    • Make sure to let the probe rest inside the solution and click ‘OK’ when stabilization icon appears on the Meter.
  2. Add the ‘Conductivity Meter Calibrate’ action and select the conductivity meter configured in Hardware Setup.
  3. Complete the script, configure the ‘Run Configuration’ table, and click ‘Go to Run’ to start the meter calibration.