What are the COD/BOD ranges for the PeCOD® and what are the mixing ratios?

This depends on the reagent range. There are 3 COD reagent ranges for the PeCOD®. The advanced blue range is the lowest range and analyzes samples up to 25mg/L with a mixing ratio of 3:1 (sample to electrolyte). Green is the second lowest range and measures up to 150mg/L with a mixing ratio of 1:1.The yellow range determines COD up to 12,000mg/L with a mixing ratio of 1:9. For more information, read MANTECH’s technical bulletin here.