Frequently Asked Questions
BOD Analyzer
General Information
What field bottles are recommended for automated sample transfer, pH adjustment, and dechlorination?
MANTECH offers custom racks to accommodate customer’s existing field bottles as well as standard options for 500mL and 1L bottles. See details below.
500mL Bottles
- Dimensions: 2.41”/6.12cm Dia. x 7.58”/19.25cm Bottle Hgt. x 0.47″/1.19cm Cap Hgt.
- Capacity per rack: 18 bottles
- View bottles here
1L Bottles
- Dimensions: 3.58”/9.1cm Dia. x 7.83”/19.9cm Bottle Hgt.
- Capacity per rack: 10 bottles
- View bottles here
It is important to keep in mind the total volume required to prepare all dilutions.
What are the dimensions of MANTECH BOD autosampler racks?
| Part Number | Rack Size | Autosampler Model | Bottle Style | Depth | Width | Rack Height | Rack Height W/ Bottles |
| PB-10246 | 18 x 3000mL | AM400 & AM300 Series | Glass | 487.68mm
19.2in |
236.22mm
9.03in |
63.5mm
2.5in |
2.69mm + bottle height
0.106in + bottle height |
| PB-10247 | 18 x 300mL | AM400 & AM300 Series | Plastic | 487.68mm
19.2in |
236.22mm
9.03in |
63.5mm
2.5in |
2.69mm + bottle height
0.106in + bottle height |
| PB-10248 | 12 x 300mL | AM400 Series | Glass | 487.68mm
19.2in |
153.42mm
6.04in |
63.5mm
2.5in |
2.69mm + bottle height
0.106in + bottle height |
| PB-10255 | 12 x 300mL | AM400 Series | Plastic | 487.68mm
19.2in |
153.42mm
6.04in |
63.5mm
2.5in |
2.69mm + bottle height
0.106in + bottle height |
| PB-10138 | 11 x 300mL | AM73 | Plastic | 337mm
13.25in |
234mm
9.19in |
88mm
3.13in |
8mm + bottle height
0.313in + bottle height |
| PB-10139 | 11 x 300mL | AM73 | Glass | 337mm
13.25in |
234mm
9.19in |
88mm
3.13in |
8mm + bottle height
0.313in + bottle height |
| PB-10162 | 23 x 300mL | AM122 | Glass | 307mm
12.06in |
307mm
12.06in |
71mm
3.81in |
9.5mm + bottle height
0.375in + bottle height |
| PB-10163 | 23 x 3000mL | AM122 | Plastic | 315mm
12.38in |
307mm
12.06in |
71mm
3.81in |
9.5mm + bottle height
0.375in + bottle height |
What incubators do MANTECH recommend?
The 5-day Biochemical Oxygen Demand (BOD5) test requires that samples be incubated for a period of 5 days between the initial reading of dissolved oxygen, and the final reading. The environment the samples are incubated in must be strictly controlled at 20°C for the entirety of the 5 days, and the best option for establishing this environment is an incubator. There are a wide variety of incubators available in the market with different dimensions and volumes, accommodating more or less samples depending on a laboratory’s requirements.
MANTECH’s BOD Pro analyzer automates the preparation and analysis of sets of samples for this BOD5 test using a multi-position autosampler along with pumps and other components. MANTECH’s AM400 autosamplers all accommodate 12- position or 18-position racks which the operator places the sample bottles into, then starts the run. After the analysis is complete, the operator applies stoppers and overcaps to the bottles according to the Standard Methods, and the sample bottles are placed in the incubator. For the BOD Pro systems, it is greatly beneficial to keep the bottles in position on the racks and place the full racks in the incubator, to save time and effort for the operators.
The dimensions of an 18-position BOD rack with bottles, stoppers and overcaps applied are provided below:
L – 19in/49cm
W – 9in/23cm
H – 7.5in/19cm
The dimensions of a 12-position BOD rack with bottles, stoppers and overcaps applied are provided below:
L – 19in/49cm
W – 6in/15cm
H – 7.5in/19cm
When sourcing an incubator for your lab’s BOD analysis, you will want to consider the total number of bottles expected to be analyzed every week, as well as the number of racks those bottles will occupy. For example, if your lab analyzes 200 BOD bottles every week, this would correspond to 12 18-position racks, with the last rack only being partially filled but still taking up the space of a full rack in the incubator.
Some examples of recommended incubator models and their bottle capacities both including racks, and not including racks, are provided below:
| Brand | Model # | Description | Capacity with 18-position Racks | Capacity with 12-position Racks | Capacity w/o Racks | Link |
| Fisher Scientific | S50441A | Economy Standard Incubator, 19.8 L, Aluminum | x | x | 12 | https://www.fishersci.ca/ |
| Fisherbrand | 15-015-2633 | Microbiological Incubator, 60 L, Stainless Steel | x | x | 75 | https://www.fishersci.ca/ |
| Fisherbrand | 15-015-2634 | Microbiological Incubator, 117 L, Stainless Steel | x | x | 108 | https://www.fishersci.ca/ |
| Fisherbrand | 15-015-2645 | Microbiological Incubator, 180 L, Stainless Steel | 108 | 108 | 144 | https://www.fishersci.ca/ |
| Fisherbrand | 15-103-1569 | Isotemp™ 400 Refrigerated Incubator, 381 L, Stainless Steel | 252 | 252 | 343 | https://www.fishersci.ca/ |
| Fisherbrand | 3720A | Isotemp™ BOD Refrigerated Incubator, 566 L, Porcelain Steel | 288 | 384 | 441 | https://www.fishersci.ca/ |
| Thermo Scientific | 13-262-340 | Forma™ Environmental Chamber Model 3920, 821.2 L, Stainless Steel | 432 | 576 | 792 | https://www.fishersci.ca/ |

PeCOD® COD/BOD Analyzer
General Information
What is the typical initial PeCOD® shipping weight and dimensions?
An investment in the safe, green, and fast PeCOD® Analyzer includes the PeCOD® L50 Analyzer, starter kits, and consumables. This is typically packed in two cartons with the following weight and dimensions: Box 1 (starter kits and consumables) – 16” x 15” x 14”, 32lbs/41cm x 39cm x 36cm, 15kg, Box 2 (PeCOD® L50) – 16” x 15” x 16”, 21lbs/41cm x 39cm x 41cm, 10kg
Other
General
General Information
What are the dimensions & weights of MANTECH systems & crates?
Please see table here.
Support and Troubleshooting
Hardware
What are the dimensions and space requirements for MANTECH systems?
MANTECH’s AutoMax (AMXXX) autosamplers and On-Line (MO) systems come in a variety of sizes to meet different space and capacity requirements. View all the dimensions and space requirements here.
Get Smarter Solutions, Today.
Our innovative water and wastewater analyzers have faster, safer processes, with fewer hazardous reagents. Start using MANTECH solutions for more accurate data that drives responsible outcomes.