Frequently Asked Questions
MT Analyzer
Applications
How are individual carbonate, bicarbonate, and hydroxide species reported in their own ionic concentrations from Alkalinity?
Once Total Alkalinity (T-Alk) and p-Alkalinity (p-Alk) are calculated, additional equations are created to back-calculate how much of the alkalinity exists as carbonate (CO3²⁻), bicarbonate (HCO3⁻), and hydroxide (OH⁻) using each species’ own equivalent weight, so the speciation is reported in true ionic concentrations rather than as CaCO3 equivalents. View more details here.
What is Gran analysis and why is Gran alkalinity important?
Gran analysis is an effective alternative method for endpoint and pKa detection. Using a series of mathematical manipulations, standard titration curves are transformed into linear data called ‘Gran functions’. Endpoints and pKa’s are determined by performing a linear regression on these functions. In many cases, Gran analysis provides a more accurate endpoint, or identifies endpoints not evaluated in Standard Methods.
How is alkalinity speciated into carbonate, bicarbonate and hydroxide fractions?
The main compounds of alkalinity are: hydroxides (OH–), carbonates (CO32-), and bicarbonates (HCO3–). The alkalinity, or buffering capacity, of a solution depends on the absorption of positively charged hydrogen ions by negatively charged bicarbonate and carbonate molecules. When bicarbonate and carbonate molecules absorb hydrogen ions, there is a shift in equilibrium without a significant shift in pH. A sample with high buffering capacity will have high bicarbonate and/or carbonate content, and a greater resistance to changes in pH.
For more information on MANTECH’s method for automated alkalinity measurement and how the species of alkalinity are calculated, download the pdf.
What is Alkalinity?
Alkalinity refers to the capability of water to neutralize acid, also known as an expression of buffering capacity. A buffer is a solution to which an acid can be added without changing the concentration of available H+ ions (without changing the pH) appreciably. In other words, a buffer absorbs the excess H+ ions and protects the water body from fluctuations in pH.
Read MANTECH’s alkalinity method abstract here
General Information
What is the minimum total liquid volume that can be measured for pH, conductivity and alkalinity with the MT Systems?
When using 50mL sample tubes, the minimum total volume can be as small as 6 ml using the TitraPro4 pH electrode and the MANTECH 5mm conductivity probe. This is due to the lower immersion depth of these electrodes and precise autosampler coordinate specifications. When using 125mL sample cups, the minimum total volume can be as small as 15mL.
What is the time of analysis per parameter and for multi-parameter analysis?
The timing for analysis of different parameters depends on several different factors, like the concentration of the sample and the equipment used.
Another factor is the total available sample volume in the vessel on the AutoMax Autosampler bed. With some combinations of parameters, larger sample volume allows for simultaneous measurements, resulting in faster total analysis time than sequential measurement. A larger sample volume may require a larger sample cup, which decreases capacity on the same model Autosampler. For example, an AM401 can accommodate 112x50ml tubes, or 62x125ml cups. In some, the cases the same set of parameters can be analyzed in 2-5 minutes faster when using the 125ml cups vs 50ml tubes.
It is important to understand your requirements in terms of capacity and throughput speed, total batch capacity required and/or daily sample loads and if overnight, unattended analyses will be applicable.
Individual Parameters
| Parameter | Average Time of Analysis (Minutes) per Sample |
| pH | 1 – 3 |
| ISE by Direct Measure | 1 – 5 depending on sample concentration |
| Titrations (Alkalinity, Acidity, Hardness) | 2 – 5 depending on sample concentration, injection rates and stability control in the titration method. |
| Conductivity | 1 – 2 |
| Colour | 2 – 3 |
| Turbidity | 2 – 3 |
Combined Parameters
MT10 & MT30 Models
| Parameters | Estimated Time of Analysis (Minutes) per Sample* | |
| 50mL Tubes | 125mL Cups | |
| pH-Alkalinity | 3 – 4 | |
| Conductivity-pH-Alkalinity | 3 – 4 | |
| Conductivity-pH-Turbidity | 4 – 7 | 3 – 4 |
| Conductivity-pH-Turbidity-Colour
(only MT30 for these suit of parameters) |
6 – 9 | 3 – 5 |
MT100 Model
| Parameters | Estimated Time of Analysis (incl. TitraSip & RapidAlk) | |
| 50mL Tubes | 125mL Cups | |
| Conductivity-pH-Alkalinity | 3 – 5 | 3 – 5 |
| Conductivity-pH-Alkalinity-Fluoride | 7 – 10 with single TitraSip or 3 – 5 with RapidDuo configuration. | |
| Conductivity-pH-Alkalinity-Turbidity | 7 – 10 | 6 – 8 |
| Conductivity-pH-Alkalinity-Colour | 4 – 6 | 4 – 6 |
| Conductivity-pH-Alkalinity-Turbidity-Colour | 8 – 11 | 4 – 6 |
Other
Support and Troubleshooting
Applications
What should I do if my titration standard is measuring too high or too low?
If your titration standards are not reading the correct concentrations, for example, the alkalinity standard reading is low, make sure the titrant has been standardized. The precision of the results can indicate if this is a mechanical or chemical problem. If the results are precise, it is likely a chemical issue. Check your standards and titrant standardization. It is also possible that the sample volume may be incorrect.
MANTECH Pro Software™
How to add gran alkalinity method to MANTECH Pro Software™?
Follow this step-by-step PDF on adding gran alkalinity method to MANTECH Pro Software™.