Frequently Asked Questions

BOD Analyzer

General Information

What incubators do MANTECH recommend?

The 5-day Biochemical Oxygen Demand (BOD5) test requires that samples be incubated for a period of 5 days between the initial reading of dissolved oxygen, and the final reading. The environment the samples are incubated in must be strictly controlled at 20°C for the entirety of the 5 days, and the best option for establishing this environment is an incubator. There are a wide variety of incubators available in the market with different dimensions and volumes, accommodating more or less samples depending on a laboratory’s requirements.

MANTECH’s BOD Pro analyzer automates the preparation and analysis of sets of samples for this BOD5 test using a multi-position autosampler along with pumps and other components. MANTECH’s AM400 autosamplers all accommodate 12- position or 18-position racks which the operator places the sample bottles into, then starts the run. After the analysis is complete, the operator applies stoppers and overcaps to the bottles according to the Standard Methods, and the sample bottles are placed in the incubator. For the BOD Pro systems, it is greatly beneficial to keep the bottles in position on the racks and place the full racks in the incubator, to save time and effort for the operators.

The dimensions of an 18-position BOD rack with bottles, stoppers and overcaps applied are provided below:

L – 19in/49cm

W – 9in/23cm

H – 7.5in/19cm

The dimensions of a 12-position BOD rack with bottles, stoppers and overcaps applied are provided below:

L – 19in/49cm

W – 6in/15cm

H – 7.5in/19cm

When sourcing an incubator for your lab’s BOD analysis, you will want to consider the total number of bottles expected to be analyzed every week, as well as the number of racks those bottles will occupy. For example, if your lab analyzes 200 BOD bottles every week, this would correspond to 12 18-position racks, with the last rack only being partially filled but still taking up the space of a full rack in the incubator.

Some examples of recommended incubator models and their bottle capacities both including racks, and not including racks, are provided below:

Brand Model # Description Capacity with 18-position Racks Capacity with 12-position Racks Capacity w/o Racks Link
Fisher Scientific S50441A Economy Standard Incubator, 19.8 L, Aluminum x x 12 https://www.fishersci.ca/
Fisherbrand 15-015-2633 Microbiological Incubator, 60 L, Stainless Steel x x 75 https://www.fishersci.ca/
Fisherbrand 15-015-2634 Microbiological Incubator, 117 L, Stainless Steel x x 108 https://www.fishersci.ca/
Fisherbrand 15-015-2645 Microbiological Incubator, 180 L, Stainless Steel 108 108 144 https://www.fishersci.ca/
Fisherbrand 15-103-1569 Isotemp™ 400 Refrigerated Incubator, 381 L, Stainless Steel 252 252 343 https://www.fishersci.ca/
Fisherbrand 3720A Isotemp™ BOD Refrigerated Incubator, 566 L, Porcelain Steel 288 384 441 https://www.fishersci.ca/
Thermo Scientific 13-262-340 Forma™ Environmental Chamber Model 3920, 821.2 L, Stainless Steel 432 576 792 https://www.fishersci.ca/

What information is required to automate biochemical oxygen demand?

Automating the biochemical oxygen demand (BOD) test can greatly increase the efficiency and production capacity of a laboratory. In order to automate this analysis, let us know how many bottles on average do you set up for an initials batch (for example, this accounts for all samples and all dilutions per sample, plus blanks, GGA’s, etc.), how often do you set up Initials (for example, 3, 5, or 7 days per week), whether or not you analyzer BOD, cBOD or both, the source of most of the samples you are analyzing (e.g., municipal WW, industrial WW, surface water, etc.), the percentage of samples you know the approximate BOD of before analysis and who your clients or users of the BOD information are (e.g., industrial pre-treatment, WWTP effluent compliance, WWTP engineers, etc.).

BOD Application

What is the difference between BOD and CBOD?

Biochemical oxygen demand (BOD) is a way to assess the amount of oxygen required for aerobic microorganisms to decompose the organic material in a sample of water over a specific time frame. It is the oxygen uptake demand of a source of water. Carbonaceous biochemical oxygen demand (CBOD) is the same method as BOD, but the nitrifying bacteria in the sample are inhibited. Nitrifying bacteria consume nitrogenous materials (compounds with reduced forms of nitrogen) and add to the oxygen demand of the wastewater. Nitrogenous materials are often seen as interference because the purpose of the BOD test is to measure carbonaceous material.

What is biochemical oxygen demand (BOD)?

Biochemical Oxygen Demand (BOD), also often referred to as biological oxygen demand, is a test performed to measure the potential of wastewater and other waters to deplete the oxygen level of receiving waters. In other words, the BOD test is performed to determine what effect dirty water, containing bacteria and organic materials, will have on animal and plant life when released into a stream or lake.

PeCOD® COD/BOD Analyzer

COD Application

What is the difference between chemical oxygen demand and biochemical oxygen demand?

Chemical Oxygen Demand (COD) analysis is a measurement of the oxygen-depletion capacity of a water sample contaminated with organic waste matter. Specifically, it measures the equivalent amount of oxygen required to chemically oxidize organic compounds in water. The traditional COD method is the wet chemistry method, dichromate COD (CODCr). This involves a two hour digestion at high heat under acidic conditions and involves hazardous chemicals such as mercury and dichromate. Biochemical Oxygen Demand (BOD), also often referred to as biological oxygen demand, is a test performed to measure the potential of wastewater and other waters to deplete the oxygen level of receiving waters. The BOD test involves taking an initial dissolved oxygen (DO) reading and a second reading after five days of incubation at 20°C. For this reason, this test is often written as BOD5 for short. MANTECH has developed an automated COD method utilizing a new, rapid and green technology called the PeCOD® COD Analyzer. This method directly measures the amount of oxidizable material in a sample via photoelectrochemical oxidation in a microcell, eliminating the need for time-consuming digestion and hazardous chemicals as only an electrolyte solution is required for analysis.

What is chemical oxygen demand (COD)?

Chemical Oxygen Demand (COD) analysis is a measurement of the oxygen-depletion capacity of a water sample contaminated with organic waste matter. Specifically, it measures the equivalent amount of oxygen required to chemically oxidize organic compounds in water. COD is used as a general indicator of water quality and is an integral part of all water quality management programs. Additionally, COD is often used to estimate BOD (Biochemical Oxygen Demand) as a strong correlation exists between COD and BOD, however COD is a much faster, more accurate test. Learn more here.

General Information

How does the PeCOD® Analyzer calculate COD?

The PeCOD® analyzer performs advanced oxidation on a small volume of sample. As the reaction proceeds, electrical charge is generated proportional to the oxygen being consumed. The PeCOD® analyzer captures this generated charge, plotting the output current from the reaction over time as shown below. The area under the curve generated by plotting current over time is proportional to the oxygen demand of the sample. A blank charge is also determined for each sample, and subtracted from the total charge to ensure accuracy.

View the detailed overview of peCOD technology and calculations here.

What information is required for PeCOD® analysis?

PeCOD® can greatly increase the efficiency and transparency of chemical oxygen demand (COD) or estimated biochemical oxygen demand (BOD) results of a laboratory, factory, wastewater treatment plant and more. To perform this analysis, the following information is required:

  1. What is the COD or BOD concentration range to measure?
    • COD:
    • BOD:
  2. How many samples will be analyzed per day or per week? Whichever applies.
  3. Why do you measure COD currently? For example, production control, effluent compliance, etc.
  4. Do you do on-site treatment? If so, What is the treatment train?

Can the PeCOD® analyzer analyze sludgy/dirty samples with large particles?

For samples containing particles larger than 50μm, a pre-filtration or homogenization step is required. For biochemical oxygen demand (BOD) analysis, a simple pre-filtration step is required. MANTECH has a Sample Filtering Guide for PeCOD® Analysis. For chemical oxygen demand (COD) analysis, no filtering required as we recommend a homogenizer that gets all particles to a 1micron size.  MANTECH recommends:

PRO200 Homogenizer with SS Probe – This is our base model and setup can be used within 5ml-50ml tubes with sample piece size of 10mm or less or liquid base sample up to 100ml. (MANTECH part number: PCH-4000-197)

PRO250 Homogenizer with SS Probe – When there is a more viscous/thick or larger solid sample piece over 10mm, our 20mm X 115mm SS Probe would be needed. This homogenizer & probe setup can cover 50ml-500ml tubes/beakers. Other probes can be used with this Homogenizer for even larger volumes. (MANTECH part number: PCH-4000-198)

Homogenization can be automated using the autosampler arm and in our online analyzer.

For pulp and paper and wastewater applications, MANTECH recommends using a 35um polyethylene (PE) syringe filter.  These filters can contribute trace amounts of organics, which are negligible for wastewater applications.  For drinking and source water applications it’s important to use a filter that does not contribute organics to the filtered sample.  One of MANTECH’s research partners has recommended a 0.45um polyether sulfone (PES) filter; however, other filter types may also be acceptable, if no organics are contributed by the filter.  Since these applications traditionally see less particulates, having a smaller pore size filter hasn’t shown an impact on the PeCOD® results.

Will PeCOD® correlate with my Dichromate COD results?

There is a strong correlation between the PeCOD COD results and the dichromate COD results. To determine this, the two methods were compared vs. the theoretical oxygen demand (ThOD) for 34 organic species. See MANTECH’s technical bulletin for more information on the study.

Does PeCOD® conform to regulatory standards?

PeCOD® is utilized as an internal process control tool to make impactful real-time decisions.The PeCOD® conforms to regulatory standards such as the ASTM International Method D8084, the Ministry of Environment, Conservation and Parks, Ontario (MECP) method E3515, and the Health Canada Guidance on drinking water. The ASTM International method for photoelectrochemical oxygen demand is approved for measuring organics in freshwater sources and treated drinking water. The Ontario MECP method E3515 replaced the standard dichromate methods due to the fact that no harmful chemicals are used in the peCOD method. This method now includes PeCOD® as an approved alternate COD method in the Municipal and Industrial Strategy for Abatement (MISA). The PECOD method is also referenced in the Health Canada Guidance on Natural Organic Matter in Drinking Water. COD has been added as a parameter with a <5ppm limit, only the peCOD method is referenced for the parameter, and peCOD is also referenced as a “parameter” to monitor in source waters for drinking water plants. Learn more here.

How does the PeCOD® method compare to BOD, TOC, and conventional COD?

BOD and COD methods differ from TOC because they measure the amount of oxygen that is depleted by organic species in water. There are several common methods to test wastewater and drinking water for organic pollutants, natural and chemical.  Chemical Oxygen Demand (COD), Biochemical Oxygen Demand (BOD) and, Total Organic Carbon (TOC) compromise the three main methods of testing water samples.  BOD and COD methods differ from TOC because they measure the amount of oxygen that is depleted by organic species in water.  Moreover, TOC is a measure of all carbon (both organic and inorganic), rather than the oxygen that is reduced by these species.  As written by a TOC manufacturer, “TOC on its own sheds no light on the oxidizability of the measured carbon or the amount of oxygen needed for its biodegradation.”  Specific to COD, it measures the reactive fraction of the TOC.  This is also known as oxidizability in the European Union. View the full article: Comparison of COD, BOD and TOC Methods for organics, which summarizes the advantages and disadvantages of the COD, BOD and TOC methods, and compares them to the PeCOD method.

How does the PeCOD® analyzer determine oxygen demand?

The PeCOD® analyzer performs advanced oxidation on a small volume of sample. As the reaction proceeds, electrical charge is generated proportional to the oxygen being consumed. The PeCOD® analyzer captures this generated charge, plotting the output current from the reaction over time as shown below. The area under the curve generated by plotting current over time is proportional to the oxygen demand of the sample. A blank charge is also determined for each sample, and subtracted from the total charge to ensure accuracy.

View the detailed overview of peCOD technology and calculations here.

TOC Analyzer

General Information

What is oxygen demand?

OD (Oxygen Demand) measures the chemical reactivity of organics by the demand for oxygen, as shown in the diagram below. It can be used as an additional tool in the characterization of NOM (Natural Organic Matter) to predict DBP (Disinfection by-product) formation. This metric also allows for rapid feedback and optimization of coagulation and disinfection dose requirements.

How does the PeCOD® method compare to BOD, TOC, and conventional COD?

BOD and COD methods differ from TOC because they measure the amount of oxygen that is depleted by organic species in water. There are several common methods to test wastewater and drinking water for organic pollutants, natural and chemical.  Chemical Oxygen Demand (COD), Biochemical Oxygen Demand (BOD) and, Total Organic Carbon (TOC) compromise the three main methods of testing water samples.  BOD and COD methods differ from TOC because they measure the amount of oxygen that is depleted by organic species in water.  Moreover, TOC is a measure of all carbon (both organic and inorganic), rather than the oxygen that is reduced by these species.  As written by a TOC manufacturer, “TOC on its own sheds no light on the oxidizability of the measured carbon or the amount of oxygen needed for its biodegradation.”  Specific to COD, it measures the reactive fraction of the TOC.  This is also known as oxidizability in the European Union. View the full article: Comparison of COD, BOD and TOC Methods for organics, which summarizes the advantages and disadvantages of the COD, BOD and TOC methods, and compares them to the PeCOD method.

Other

Support and Troubleshooting

PeCOD®

What are the reasons for differences between the PeCOD® COD and Dichromate COD results in some cases?

Chemical Oxygen Demand (COD) results may differ when measured via the PeCOD® COD method versus the traditional dichromate COD method for certain sample matrices. There are various reasons for this difference. One is that chloride, ammonia, and some heavy metals have been known to interfere with PeCOD® readings and provide inaccurate results. Another reason could be the time delays between analyses. It is best to analyze samples via PeCOD® and dichromate on the same day to limit uncertainties due to sample degradation.

For additional reasons and more information, read MANTECH’s technical bulletin here.