Frequently Asked Questions

BOD Analyzer

General Information

How long does a batch take for Automated Sample Transfer, pH Adjustment, & Dechlorination + Automated BOD Preparation & Uncapping/Capping system?

MANTECH Pro™ prepares and transfers all samples first (pH adjustment, dechlorination, transfer), then BOD Pro™ fills the bottles and reads the initial DOs. For 18 field bottles with 3 dilutions each, expect about 3.0 to 3.3 hours with Buret seed when every sample needs pH adjustment and dechlorination, or about 2.4 hours when neither is needed. View a full breakdown of the timing and other examples here.

What field bottles are recommended for automated sample transfer, pH adjustment, and dechlorination?

MANTECH offers custom racks to accommodate customer’s existing field bottles as well as standard options for 500mL and 1L bottles. See details below.

500mL Bottles

  • Dimensions: 2.41”/6.12cm Dia. x 7.58”/19.25cm Bottle Hgt. x 0.47″/1.19cm Cap Hgt.
  • Capacity per rack: 18 bottles
  • View bottles here

1L Bottles

  • Dimensions: 3.58”/9.1cm Dia. x 7.83”/19.9cm Bottle Hgt.
  • Capacity per rack: 10 bottles
  • View bottles here

It is important to keep in mind the total volume required to prepare all dilutions.

Can MANTECH’s systems automate pH adjustment?

Yes, MANTECH offers benchtop and autosampler options to automate pH measurement and adjustment according to Standard Methods.

BOD Application

Are BOD results affected by atmospheric pressure changes?

Generally speaking, dissolved oxygen readings are dependent on atmospheric pressure. This is because the density of water changes with pressure, and this in turn affects the mg/L of dissolved oxygen. Here is a resource describing the “why” behind this. It is also dependent on temperature for similar reasons.The YSI meter does compensate for changes in atmospheric pressure. Here is a document explaining this. The document also provides advice for troubleshooting when challenges are occurring in BOD testing. One important aspect is to make sure you are calibrating the probe before each day of testing.

MT Analyzer

Applications

How can the Permanganate Index test be automated?

MANTECH is able to automate all aspects of the permanganate index test including automated sample heating, acid and reagent addition, and titration with calculation of results. The analysis can also be optimized for maximum throughput capacity through a batch setup where multiple samples at once are heated, prepared, and analyzed all within the timing specified in the ISO 8467 method. Overall, the benefit of the batching process is much quicker time per sample overall since you are making use of the 10 minute wait times between adding the reagents instead of the system just sitting and waiting.

 

The time per sample for single analysis is roughly 25 minutes per sample taking into account all the waiting times and about 3 minutes for the actual titration.

 

For the batch analysis, you get roughly 12 samples per batch and the batch takes roughly 60 minutes to complete, making it only about 5 minutes per sample.

 

Below are the steps both for single analysis and batch analysis:

 

Single Analysis:

  1. A blank analysis performing the steps 3-11 listed below, using DI water as the sample, is run and the blank endpoint volume is used in the calculation of sample results for the run.
  2. After titrating the blank sample, 5mL of 0.005M Sodium Oxalate is added to the titrated sample, then after a brief mixing period the titration is performed again. The endpoint volume is recorded as the standardization volume and is also used in the calculation of sample results for the run.
  3. 25mL of the sample volume is pipetted into each sample beaker, and sample protect caps are placed on the beakers.
  4. The Heating Element are turned on. Heating is to be applied keeping the sample temperature between 96-98°C during the following steps until the element is turned off.
  5. The Sample Protect cap is removed and 5mL of 2.0M Sulfuric Acid is pipetted into a beaker. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed on the beaker.
  6. 10-minute wait time.
  7. The Sample Protect cap is removed and 5mL of 0.002M Potassium Permanganate solution is added. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed back on the beaker.
  8. 10-minute wait time.
  9. The Sample Protect cap is removed and 5mL of 0.005M Sodium Oxalate solution is added. The stirrer turns on to mix, at which point the solution will briefly change color but eventually become colorless.
  10. 1-minute wait time.
  11. The solution is titrated with 0.002M Potassium Permanganate to a faint pink color. This color is measured with the photometric probe.
  12. Steps 5 to 11 are repeated for all remaining samples.

 

Batch Analysis:

  1. A blank analysis performing the steps 3-13 listed below, using DI water as the sample, is run and the blank endpoint volume is used in the calculation of sample results for the run.
  2. After titrating the blank sample, 5mL of 0.005M Sodium Oxalate is added to the titrated sample, then after a brief mixing period the titration is performed again. The endpoint volume is recorded as the standardization volume and is also used in the calculation of sample results for the run.
  3. 25mL of the sample volume is pipetted into each sample beaker, and sample protect caps are placed on the beakers.
  4. The Heating Element are turned on. Heating is to be applied keeping the sample temperature between 96-98°C during the following steps until the element is turned off.
  5. The Sample Protect cap is removed and 5mL of 2.0M Sulfuric Acid is pipetted into a beaker. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed on the beaker.
  6. Step 5 is repeated for multiple sample beakers in a row, until 10 minutes has elapsed since the acid was added to the first beaker. The set of beakers that get acid added to them within the 10 minute time become the first “batch”.
  7. After 10 minutes has elapsed, the system moved back to the first sample beaker in the batch.
  8. The Sample Protect cap is removed and 5mL of 0.002M Potassium Permanganate solution is added. The stirrer turns on briefly to mix the solution, then the sample protect cap is placed back on the beaker.
  9. Step 8 is repeated for the remaining beakers in the batch determined from Step 6. Roughly 10 minutes should elapse from when the permanganate was added to the first beaker.
  10. After 10 minutes has elapsed, the system moved back to the first sample beaker in the batch.
  11. The Sample Protect cap is removed and 5mL of 0.005M Sodium Oxalate solution is added. The stirrer turns on to mix, at which point the solution will briefly change color but eventually become colorless.
  12. 1-minute wait time.
  13. The solution is titrated with 0.002M Potassium Permanganate to a faint pink color. This color is measured with the photometric probe.
  14. Steps 11-13 are repeated for the remaining samples in the first batch.
  15. After the first batch is complete, the system repeats steps 5-14 for all remaining samples in the run, creating as many “batches” as required to complete the full set of samples.

Watch MANTECH’s Automated Permanganate Index System in action here

What methods of lime analysis can be automated?

MANTECH systems conform to Tex-600-J and Tex-406-A Part III. Automated titrations (and back titrations) utilizing hydrochloric acid, sodium hydroxide, and other solutions are automated via MANTECH’s high precision 100,000 step buret with pH measuring range from 0-14.  Tex-600-J includes sampling and testing of: hydrated lime, quicklime, commercial lime slurry and carbide lime slurry.

How can I determine true or filtered colour using turbidity measurement instead of filtering of the sample?

True or Filtered Colour is determined by filtering samples through a 0.45 micron filter before analyzing the sample in a spectrophotometer. When running large amounts of samples on automated systems, this filtering can become time consuming and heavy on resource usage. An alternative method for determining true colour without filtering has been established, through the use of a turbidity meter to quantify the contribution from solids present in the sample. MANTECH can easily combine automated Turbidity and Colour analysis on one system, determined from one sample, to provide True Colour results without the need to filter.

The determination is as follows:

true-color-equation

  • C is the colour result.
  • A is the sample absorbance.
  • l is the cell path length.
  • E is the scattering coefficient estimated at 0.00264 NTU-1 cm-1.
  • t is the turbidity.
  • e is the color absorptivity coefficient.

Why do I need to standardize my NaOH?

NaOH is highly hygroscopic, meaning that it absorbs water from the air. Therefore, over time the titrant will become more dilute as it absorbs water. NaOH can be standardized by titrating into a sample of potassium hydrogen phthalate (KHP) of known concentration. For example, titrate 0.05 N KHP with 0.1 N NaOH to an endpoint, and using the volume of NaOH added, the precise concentration of NaOH can be calculated. For more information on standardizing NaOH titrant, please refer to Standard Methods 2310. To limit the amount of water being absorbed, a glass drying tube with a cotton ball inserted should be used to prevent moisture going into the tube.

What is the method for hot acidity determination?

Hot acidity is a variation of the standard Acidity titration using NaOH, with some pretreatment steps performed before the titration. It is outlined in Standard Methods SM2310B, officially called “hot peroxide treatment procedure for acidity determination”. This method is applicable to mine wastes and other samples containing large amounts of metals such as heavy industrial WW.

From SM2310B:

  • Use the hot peroxide procedure to pretreat samples known or suspected to contain hydrolysable metal ions or reduced forms of polyvalent cation, such as iron pickle liquors, acid mine drainage, and other industrial wastes.
  • Hot peroxide treatment acidity method:
    • Pipet sample volume into vessel
    • Measure pH
    • If pH > 4.0, add small increments of 0.02N H2SO4 to reduce pH to 4.0 or less
    • Remove samples from MANTECH system
    • Add 5 drops of 30% H2O2 and boil for 2 to 5 minutes
    • Cool to room temperature
    • Place samples back on MANTECH system and titrate with standard alkali (NaOH) to pH 8.3
  • The calculation for hot acidity is shown below:
    • Acidity = [((mL NaOH consumed)*(Normality of NaOH)) – ((mL H2SO4 consumed)*(Normality of H2SO4))*50,000]/Sample volume

What are the different types of calibration profiles?

Calibrations can be linear, logarithmic, single-line, and multi-line fit. The calibration method used depends on the method required. pH and colour calibrations are linear, single-line fit calibrations. Conductivity and turbidity calibrations are multi-line, linear type calibrations. ISE (fluoride, chloride, ammonia, etc.) use a logarithmic multi-line calibration type.

What is the recommended type standard for automated Turbidity Calibrations and Quality Control?

MANTECH recommends the use of turbidity standards made with suspensions of microspheres of styrene-divinylbenzene copolymer for all turbidity applications. Standard Methods dictate that “Secondary standards made with suspensions of microspheres of styrene-divinylbenzene copolymer typically are as stable as concentrated formazin and are much more stable than diluted formazin.” Any dilutions used to prepare turbidity standards, as well as blanks, must use degassed DI water. This is especially important for accurate low-end turbidity analysis, as there is often a positive bias due to bubbles introduced into the DI water solution when it is being produced/dispensed. This effect becomes more noticeable with lower NTU readings, and can cause challenges with performing an accurate calibration. To degas DI water, it must be dispensed into a very clean bottle or container, and left to sit for ~24h. You will see bubbles form on the side of the container as the water degasses. After degassing, it is ready to use for blanks and standards preparation.

What are the wavelengths used by the T10 turbidity meter?

The IR (infrared) light version of the T10 and T20 turbidity meters use an emitter and detector specifically at 860nm. The WL (white light) version of the meter uses a tungsten lamp and a photodiode detector, both of which operate within the full visible light spectrum. The photodiode detector has a spectral peak response between 400-600nm.

What is the wavelength measured for turbidity by Standard Methods (white light)?

Standard Method 2130B (Nephelometric Method) specifies that a laboratory or process nephelometer should have a detector system with a spectral peak response of 400 to 600 nm. MANTECH’s T10 and T20 Turbidity meters and automated turbidity applications conform to this requirement. The T10 and T20 turbidity meters are designed specifically to conform to the international requirements for reporting turbidity measurements. The IR version conforms to ISO 7027 and EN27027. The WL version conforms to APHA SM 2130B, ASTM D1889, and EPA 180.1. Both can be used on the same system if  conforming to multiple standards is required.

What is Turbidity?

Turbidity is defined as the amount of suspended particles in a solution, measured in nephelometric turbidity units, (NTU). It is used as a general indicator of the quality of water, along with colour and odour. The US EPA has a maximum contaminant level (MCL) of 5 NTU for drinking and wastewaters.

Read MANTECH’s turbidity method abstract here.

How is alkalinity speciated into carbonate, bicarbonate and hydroxide fractions?

The main compounds of alkalinity are: hydroxides (OH–), carbonates (CO32-), and bicarbonates (HCO3–).  The alkalinity, or buffering capacity, of a solution depends on the absorption of positively charged hydrogen ions by negatively charged bicarbonate and carbonate molecules.  When bicarbonate and carbonate molecules absorb hydrogen ions, there is a shift in equilibrium without a significant shift in pH.  A sample with high buffering capacity will have high bicarbonate and/or carbonate content, and a greater resistance to changes in pH.

For more information on MANTECH’s method for automated alkalinity measurement and how the species of alkalinity are calculated, download the pdf.

What is Alkalinity?

Alkalinity refers to the capability of water to neutralize acid, also known as an expression of buffering capacity. A buffer is a solution to which an acid can be added without changing the concentration of available H+ ions (without changing the pH) appreciably. In other words, a buffer absorbs the excess H+ ions and protects the water body from fluctuations in pH.

Read MANTECH’s alkalinity method abstract here

What is pH?

The potential of hydrogen (pH) is a measure of hydrogen ion (H+) concentration. Solutions with a high concentration of H+ ions have a low pH (acidic). Solutions with a low concentration of H+ ions have a high pH (alkaline).

Read MANTECH’s pH method abstract here.

How is a Method Detection Limit (MDL) defined?

A Method Detection Limit (MDL) is defined in slightly differing ways by the US EPA, and the APHA. MDL as per US EPA: The method detection limit (MDL) is defined as the minimum measured concentration of a substance that can be reported with 99% confidence that the measured concentration is distinguishable from method blank results. MDL as per APHA: (MDL) is defined as the constituent concentration that, when processed through the entire method, produces signal that has a 99% probability of being different from the blank.

General Information

How does MANTECH perform automated in-line filtering of samples?

The MANTECH color system can be set up with a multiple in line filters in parallel between the autosampler and the spectrophotometer. The system pumps sample through the filter and automatically switches from one filter to the next at the specified times. This can either be determined by a set number of samples per filter, or by a flow detection system where there is a sensor detecting when the flow through the filter drops below a certain threshold, at which point the system automatically switches to the next fresh filter and continues the run.

A general example of the run sequence using flow detection would be:

  • Sample 1 runs through filter 1, flow does not decrease below the threshold, no change of filter
  • Sample 2 runs through filter 1, flow does not decrease below the threshold, no change of filter
  • Sample 3 runs through filter 1, flow decreases below threshold for clogged valve, system switches to filter 2
  • Sample 4 runs through filter 2, flow does not decrease below the threshold, no change of filter
  • Sample 5 runs through filter 2, flow decreases below threshold for clogged valve, system switches to filter 3
  • Sample 6 runs through filter 3, flow does not decrease below the threshold, no change of filter
  • Etc. for remainder of samples, or until all filters are clogged up
    • If all filters become clogged up, run pauses and informs user to switch the filters and start back up

How does Dynamic Injection Control work on a MANTECH automated titration system?

MANTECH utilizes a method called Dynamic Injection Control while performing an automated titration to ensure that the titration curve is smooth and accurate, leading to the best precision on results. Dynamic injection control works by providing a user set volume range for each injection, with the range being between 25mL and 0.001mL for any single injection and then applying different volume injections to try and achieve a target change in the electrode measurement. The electrode measurement can either be pH units or millivolts.

For example, the titration can be set with a volume range of 0.01mL to 1.00mL, with a target change in electrode measurement of 10mV per injection. The system will start with an injection in the middle of the range, for example 0.1mL, and then look at the change in mV as a result of that injection. If the change in mV is less than the target, then the next injection increases in size, for example to 0.2mL. The system will again look at the change in mV from that last injection, then take that into account in determining the size of the next injection. Using this method, the automated titration can move quickly with large injections through portions of the titration where little change in the measurement is occurring, then automatically detect when the titration is approaching an endpoint and dynamically decrease the injections as the rate of change increases to ensure the most precise detection of the endpoint volume.

In addition to looking at measurement the rate of change, or first derivative, of the titration curve, the MANTECH automated titration system also looks at the second derivative of the curve. The second derivative represents the acceleration of the measurement and has a significant impact on the system’s ability to detect a rapid inflection, for example due to strong acid/base titrations and colorimetric titrations where the endpoint change occurs very rapidly.

The MANTECH Pro software provides the user with full control over the dynamic injection control settings, including the min/max injection volume, starting injection volume, target change in mV or pH, and other useful settings such as endpoint windows and differing smoothing settings for the first and second derivative determination. Contact MANTECH at [email protected] for more information!

What is the chemical compatibility of the tubing in MANTECH pumps?

The tubing used in MANTECH peristaltic and metering pumps is compatible with many types of chemicals. The following is a list of chemicals that are not recommended to be used with MANTECH tubing/pumps.

Chemicals without listed concentrations are assumed to be 100% concentrations/the maximum percent solubility in water.

Acetaldehyde Ethyl Benzoate Methyl Methacrylate
Acetone Ethylene Bromide Mineral Oil
Aliphatic Hydrocarbons Fluoboric Acid, 48% in w Mineral Spirits
Amyl Alcohol Fluorine Gas Motor Oil
Aqua Regia Formaldehyde, 37% in w Naphtha
Aromatic Hydrocarbons Fuel Oil Naphthalene
ASTM Reference No. 2 Oil Furfural Nitric Acid, 68-71% in w
ASTM Reference No. 3 Oil Gasoline, Automotive Nitrobenzene
Benzaldehyde Heptane Nitromethane
Benzene Hexane Oils, Essential
Benzenesulfonic Acid Hydrobromic Acid, 20 -50% in w Oils, Hydraulic (Phosphate Ester)
Bromine, Anhydrous Liquid Hydrobromic Acid, 100% in w Oils, Hydrocarbon
Butyl Alcohol Hydrofluoric Acid, 10% in w Ortho Dichlorobenzene
Carbon Disulfide Hydrofluoric Acid, 25% in w Paraffins
Carbon Tetrachloride Hydrofluoric Acid, 40 -48% in w Picric Acid, 1% in w
Chlorine, Wet Gas Isooctane Skydrol 500A
Chlorobenzene, Mono, Di, Tri Jet Fuel, JP8 Styrene Monomer
Chloroform Kerosene Sulfur Chloride
Chlorosulfonic Acid Ketones Sulfuric Acid, 95-98% in w
Cresol (m, o, or p) Lemon Oil Tetrahydrofuran
Cyclohexane Limonene-D Toluene
Cyclohexanone Lubricating Oils, Petroleum Trichloroethylene
Diesel Fuel Methyl Ethyl Ketone (MEK) Turpentine
Dioxane Methyl Isobutyl Ketone Xylene

How does MANTECH account for temperature compensation and correction in pH measurements?

As the temperature of a solution changes, the actual pH changes. This is not an error of the probe or meter being used, but is the actual pH of the solution at that particular temperature. The temperature effect on the pH value is 0.003 pH units per oC away from 25oC, per pH units away from pH 7. This effect can be either negative or positive, depending on if the temperature is above or below 25oC, and if the pH is above or below pH 7. At 25oC and pH 7, there is no change in the pH value.

The chart below shows how the actual pH changes with temperature and pH, which allows you to correct the pH reading to 25oC. As an example, if a sample measured pH 5 at a temperature of 5oC, the chart indicates this a negative effect, therefore the sample pH would be 5 – 0.12 = 4.88pH corrected to 25oC:

temperature-effect-on-pH-chart

MANTECH software accounts for temperature by recording the temperature during calibration via a thermistor probe, then it corrects the pH reading at the time of measurement to account for the difference in measured temperature of the sample vs. the calibration buffers. The reported pH value is the corrected value @ the sample temperature. The equation below shows how the PC-Titrate software calculates pH:

 

pH-equation

EMeas  =  Voltage measure by electrode at the time of titration (mV)

EInt  = Voltage of the Intercept value calculated from the calibration equation. (mV)

Slope = slope of the line calculated from the calibration equation. (mV)

TMeas = temperature measured at the time of pH measurement (K degrees)

TCal = temperature taken at the time of last calibration. (K degrees)

 

MANTECH also offers the option to report pH values corrected to a set temperature, such as 25°C. This is implemented in the software as an additional step after the sample pH is recorded. MANTECH offers this as a feature for new MT-Series systems, or as an upgrade to existing MANTECH systems.

How do MANTECH systems measure and record temperature?

The standard method for measuring temperature is with a USB-connected stainless steel thermistor probe. The specifications for this probe are below:

  • Communication: USB controlled by MANTECH Pro Software
  • Sheath Material: Super OMEGACLAD XL/ Stainless Steel/Inconel
  • Temp Range: 0 to 1038 Degree C.
  • Diameter: 0.062in
  • Length: 6 inch
  • Measurement to 3 decimal places
  • Autocalibration feature in MANTECH Pro software

Alternatively, when measuring conductivity, one can measure the temperature directly from the conductivity probe. This temperature is also displayed directly on the conductivity meter screen.

MANTECH has the ability to use either of these measurement methods for pH temperature compensation, depending on customer preference and system configuration.

MANTECH temperature sensors are available in both PT1000 and 10kNTC styles. PT1000 sensors have a linear positive-slope relationship between resistance and temperature with a resistance of 1000 ohms at 0°C, and 10kNTC sensors have a curved negative-slope relationship between resistance and temperature with a resistance of 10,000 ohms at 25°C. Both these models of temperature sensor perform well and comparably within the 0-100°C temperature range that liquid samples exist in. If you require a specific type of temperature sensor, please feel free to let MANTECH know and we will accommodate.

What is the minimum total liquid volume that can be measured for pH, conductivity and alkalinity with the MT Systems?

When using 50mL sample tubes, the minimum total volume can be as small as 6 ml using the TitraPro4 pH electrode and the MANTECH 5mm conductivity probe. This is due to the lower immersion depth of these electrodes and precise autosampler coordinate specifications. When using 125mL sample cups, the minimum total volume can be as small as 15mL.

What is the time of analysis per parameter and for multi-parameter analysis?

The timing for analysis of different parameters depends on several different factors, like the concentration of the sample and the equipment used.

Another factor is the total available sample volume in the vessel on the AutoMax Autosampler bed. With some combinations of parameters, larger sample volume allows for simultaneous measurements, resulting in faster total analysis time than sequential measurement.  A larger sample volume may require a larger sample cup, which decreases capacity on the same model Autosampler.  For example, an AM401 can accommodate 112x50ml tubes, or 62x125ml cups.  In some, the cases the same set of parameters can be analyzed in 2-5 minutes faster when using the 125ml cups vs 50ml tubes.

It is important to understand your requirements in terms of capacity and throughput speed, total batch capacity required and/or daily sample loads and if overnight, unattended analyses will be applicable.

Individual Parameters

Parameter Average Time of Analysis (Minutes) per Sample
pH 1 – 3
ISE by Direct Measure 1 – 5 depending on sample concentration
Titrations (Alkalinity, Acidity, Hardness) 2 – 5 depending on sample concentration, injection rates and stability control in the titration method.
Conductivity 1 – 2
Colour 2 – 3
Turbidity 2 – 3

Combined Parameters

MT10 & MT30 Models

Parameters Estimated Time of Analysis (Minutes) per Sample*
50mL Tubes 125mL Cups
pH-Alkalinity 3 – 4
Conductivity-pH-Alkalinity 3 – 4
Conductivity-pH-Turbidity 4 – 7 3 – 4
Conductivity-pH-Turbidity-Colour

(only MT30 for these suit of parameters)

6 – 9 3 – 5

MT100 Model

Parameters Estimated Time of Analysis (incl. TitraSip & RapidAlk)
50mL Tubes 125mL Cups
Conductivity-pH-Alkalinity 3 – 5 3 – 5
Conductivity-pH-Alkalinity-Fluoride 7 – 10 with single TitraSip or 3 – 5 with RapidDuo configuration.
Conductivity-pH-Alkalinity-Turbidity 7 – 10 6 – 8
Conductivity-pH-Alkalinity-Colour 4 – 6 4 – 6
Conductivity-pH-Alkalinity-Turbidity-Colour 8 – 11 4 – 6

 

PeCOD® COD/BOD Analyzer

COD Application

How should COD samples be stored?

MANTECH recommends storing samples in the laboratory fridge vs preserving samples. If you do choose to preserve your samples, use H2SO4 to adjust the pH of your samples to ~2.

What is the difference between chemical oxygen demand and biochemical oxygen demand?

Chemical Oxygen Demand (COD) analysis is a measurement of the oxygen-depletion capacity of a water sample contaminated with organic waste matter. Specifically, it measures the equivalent amount of oxygen required to chemically oxidize organic compounds in water. The traditional COD method is the wet chemistry method, dichromate COD (CODCr). This involves a two hour digestion at high heat under acidic conditions and involves hazardous chemicals such as mercury and dichromate. Biochemical Oxygen Demand (BOD), also often referred to as biological oxygen demand, is a test performed to measure the potential of wastewater and other waters to deplete the oxygen level of receiving waters. The BOD test involves taking an initial dissolved oxygen (DO) reading and a second reading after five days of incubation at 20°C. For this reason, this test is often written as BOD5 for short. MANTECH has developed an automated COD method utilizing a new, rapid and green technology called the PeCOD® COD Analyzer. This method directly measures the amount of oxidizable material in a sample via photoelectrochemical oxidation in a microcell, eliminating the need for time-consuming digestion and hazardous chemicals as only an electrolyte solution is required for analysis.

General Information

Why are my PeCOD values reading low in comparison to the dichromate method?

The most common cause is chloride interference.

Chloride interference can impact COD outputs no matter what methodology is used. For the PeCOD method, the interference is negative – the samples are measured as having lower COD values than expected (see the figure below for recommended limits). Meanwhile for the dichromate method, the interference is positive and results in higher COD values. This results in a significant discrepancy between the two different methods with neither measurement accurate when a chloride interference exists!

The solution for this challenge is as simple as diluting the problematic samples until the chloride concentration falls below the maximum allowance. The lowest range on the PeCOD has a maximum allowance of 400 mg/L while the dichromate method’s is 100 mg/L according to Standard Method 5220. This can be increased to 1500 mg/L with supplementary mercury sulphate. Additionally, the detection limit of COD in the PeCOD is 0.7 mg/L in Advanced Blue and 1.0 mg/L in Green range. It is 50 mg/L for the dichromate method when looking for quantitative results or 1.0 mg/L when estimated. This makes the PeCOD more suitable for both low-COD high-chloride samples as well as samples that need significant dilution due to high chloride levels.

In addition to the PeCOD method having higher chloride tolerance with lower detection limits, COD (and BOD) results are generated in 10 minutes utilizing a safe and green method.

View PDF here.

How is the PeCOD® Analyzer calibrated?

Calibrations are comprised of 6 different stages. The first stage is called the “Normalization Phase” and lasts 100 seconds. During this time, the LED strength is adjusted to maintain a baseline electrical current of 20 mA. This is performed on the blank solution, containing range specific electrolyte and COD free deionized water. Once the LED is set, oxidation of the blank solution will occur. This is comprised of 3 stages, visualized by 3 distinct curves. The stages are known as the Burn-In, Pre-Burn, and Oxidation of port B. The blank acts as a zero reference for the calibration based on the charge generated from the DI and electrolyte mixture. The area under the curve is used to quantify charge. Once the Port B stages have completed, the Port A stages will begin. The calibrant and electrolyte mixture is introduced across the TiO2 sensor, where similar Pre-Burn and Oxidation curves are generated. The concentration of the calibrant is determined based on the sample COD range. Once the specified number of calibrations have completed, the calibrant solution can be run as a sample (referred to as a QC check). It is expected that the COD result will be + / – 5% of the standard COD value. 

Will PeCOD® correlate with my BOD results?

When compared to the standard BOD test, the PeCOD® can estimate BOD concentration within a 95% confidence level.

The PeCOD® measures COD through a rapid 10-minute photoelectrochemical oxidation, allowing for the accurate monitoring of a wide range of concentrations in real-time. These COD readings can be used to reliably estimate BOD by applying a correlation coefficient. When compared to the standard BOD test, the PeCOD® can estimate BOD concentration within a 95% confidence level.

Some examples of correlation coefficients of PeCOD®/BOD that have been determined for different types of industrial wastewater are provided in the table below.

Industry PeCOD®/BOD Ratio
Beverage Company 0.56
Confectionary 0.59
Garment Industry 0.59
Packaging Company 0.63
Oil Recycling Company 0.65
Portable Sanitation 0.68
Dairy 0.91
Nut Company 0.94
Food Manufacturing 0.98
Bakery 1.52
Condensed Milk 2.15

What pH range can the peCOD method measure in?

The peCOD method requires that the pH of a sample AFTER being mixed with electrolyte must be between 4 – 10. To determine if a sample must be pH-adjusted, mix the sample with PeCOD® electrolyte at the proper mixing ratio for your COD range, then test the pH of the mixture. For example, the sample may have a pH of 3.0, but then after preparing with electrolyte, the pH is in the required range, therefore, it is acceptable for immediate PeCOD® measurement. If samples have been preserved in acid, they should be neutralized using sodium hydroxide prior to analysis to avoid a low reading, as well as damage to the sensor. When the sample pH is below 4, the photocatalytic oxidation at the TiO2 sensor is affected, leading to poor reproducibility and charge values below theoretical expectation. Below a pH of 2, the TiO2 displays instability. When the pH is above 10, the charge measured for the reference and sample solution yield lower than expected values, again caused by interference at the TiO2 sensor. Sulphuric acid should be used to lower the pH of samples with a pH of 10 or more.

Does PeCOD® conform to regulatory standards?

PeCOD® is utilized as an internal process control tool to make impactful real-time decisions.The PeCOD® conforms to regulatory standards such as the ASTM International Method D8084, the Ministry of Environment, Conservation and Parks, Ontario (MECP) method E3515, and the Health Canada Guidance on drinking water. The ASTM International method for photoelectrochemical oxygen demand is approved for measuring organics in freshwater sources and treated drinking water. The Ontario MECP method E3515 replaced the standard dichromate methods due to the fact that no harmful chemicals are used in the peCOD method. This method now includes PeCOD® as an approved alternate COD method in the Municipal and Industrial Strategy for Abatement (MISA). The PECOD method is also referenced in the Health Canada Guidance on Natural Organic Matter in Drinking Water. COD has been added as a parameter with a <5ppm limit, only the peCOD method is referenced for the parameter, and peCOD is also referenced as a “parameter” to monitor in source waters for drinking water plants. Learn more here.

TOC Analyzer

General Information

What is the primary mechanism of PeCOD® chemistry?

The primary driver of the peCOD method chemistry is advanced oxidation induced by photocatalysis with Titanium Dioxide (TiO2). Pure TiO2 is only photo-active at wavelengths below 380 nm. This is because a certain amount of light energy is required to bump the electrons around and cause the behaviors that we associate with photocatalysis. The UV LED in the PeCOD® COD Analyzer operates at a peak wavelength of 365 nm, with a minimum 360 nm and maximum 370 nm, ensuring that efficient photocatalysis is achieved.

Other

General

General Information

What is MANTECH’s mailing address?

Our corporate address is 5473 Hwy 6 North, Guelph, ON N1H 6J2 Canada. Our USA address is 22628 Island Pines Way, Unit 1604, Fort Myers Beach, FL 33931, USA.

Warranty Policies

What is MANTECH’s warranty policy?

MANTECH Instruments are warrantied for a period of one year (unless otherwise specified) from the date of shipment. This warranty covers defects in material and workmanship (parts and labour), under normal installation, use and maintenance, within the region of installation, as described in the operator’s instructions. Please refer to the user’s manual. This warranty does not cover Computer products, YSI branded products, consumables, expendable items or items with physical damage. Computer warranty/service is handled direct by the end users through the manufacturer. YSI warranty/service is handled direct by the end users through the local YSI distributor. For more information about extended warranty options, click here.

Service Agreements

Can I get preventative maintenance services without a service agreement?

Yes, preventative maintenance services can be purchased without a service agreement. There is no additional fee for a service agreement – only additional benefits including:

  • 5% discount on PM service parts
  • 15% discount on additional serviceable parts during scheduled PM visit
  • 15% discount emergency visits, in-house service and loaner modules
  • Complete evaluation of methods, as required
  • Perform software maintenance, as required
  • Additional training if required (time permitting)
  • Toll-free telephone and web-based support

What is included with a service agreement?

In addition to an annual preventative maintenance visit (on-site or remote), the following is included with a service agreement:

  • 5% discount on PM service parts
  • 15% discount on additional serviceable parts during scheduled PM visit
  • 15% discount emergency visits, in-house service and loaner modules
  • Complete evaluation of methods, as required
  • Perform software maintenance, as required
  • Additional training if required (time permitting)
  • Toll-free telephone and web-based support

Support and Troubleshooting

Applications

Why is the pH electrode slope and measuring characteristics different for higher pH values, for ex pH 13?

Changes to slope at higher pHs

Alkaline Error or Sodium Error occurs when pH is very high (e.g. pH 12) because Na+ concentration is high (from NaOH used to raise pH) and H+ is very low.

Electrodes respond slightly to Na+ and give a false low reading.  This is related to the concept of selectivity coefficients where the electrode responds to many ions but is most selective for H+.  This problem occurs because Na+ is 10 orders of magnitude higher than H+ in the solution.

High pH electrodes use a 0-14 pH glass.  This electrode will read pH 14 (1 M NaOH) to be around pH 13.7 with a 0.3 pH sodium error.

A standard pH electrode uses a 0-12 pH glass.  The electrode will read pH 14 (1 M NaOH) to be around pH 12.4 with a 1.6 pH sodium error.

 

Alkaline error

The alkaline effect is the phenomenon where H+ ions in the gel layer of the pH-sensitive membrane are partly or completely replaced by alkali ions. This leads to a pH measurement which is too low in comparison with the number of H+ ions in the sample. Under extreme conditions where the H+ ion activity can be neglected the glass membrane only responds to sodium ions.  Even though the effect is called the alkaline error, it is only sodium or lithium ions which cause considerable disturbances. The effect increases with increasing temperature and pH value (pH > 9), and can be minimized by using a special pH membrane glass.

 

Sodium Ion Error

Although the pH glass measuring electrode responds very selectively to H+ ions, there is a small interference caused by similar ions such as lithium, sodium, and potassium. The amount of this interference decreases with increasing ion size. Since lithium ions are normally not in solutions, and potassium ions cause very little interference, Na+ ions present the most significant interference.

Sodium ion error, also referred to as alkaline error, is the result of alkali ions, particularly Na+ ions, penetrating the glass electrode silicon‐oxygen molecular structure and creating a potential difference between the outer and inner surfaces of the electrode. H+ ions are replaced with Na+ ions, decreasing the H+  ion activity, thereby artificially suppressing the true pH value. This is the reason pH is sometimes referred to as a measure of the H+ ion activity and not H+ ion concentration.

Na+ ion interference occurs when the H+ ion concentration is very low and the Na+ ion concentration is very high. Temperature also directly affects this error. As the temperature of the process increases, so does the Na+ ion error.

Depending on the exact glass formulation, Na+ ion interference may take effect at a higher or lower pH. There is no glass formulation currently available that has zero Na+ ion error. Since some error will always exist, it is important that the error be consistent and repeatable. With many glass formulations, this is not possible since the electrode becomes sensitized to the environment it was exposed to prior to experiencing high pH levels. For example, the exact point at which the Na+ ion error of an electrode occurs may be 11.50 pH, after immersion in tap water, but 12.50 pH after immersion in an alkaline solution.

Controlled molecular etching of special glass formulations can keep Na+ error consistent and repeatable.

This is accomplished by stripping away one molecular layer at a time. This special characteristic provides a consistent amount of lithium ions available for exchange with the hydrogen ions to produce a similar millivolt potential for a similar condition.

Hardware

How do I store my MANTECH Electrodes?

MANTECH recommends storing pH and other electrodes immersed in a 1:10 dilution of pH 4 buffer in Tap Water. Conductivity electrodes can be stored in air for system configurations that do not place it on the same holder as another electrode.

Why is there liquid leaking from my TitraSip™ system?

The presence of liquid beneath the TitraSip™ would be related to one of two things:

  • Overflow from the vessel
  • Leaks from the tubing

Generally speaking, the rinse pump for the TitraSip™ should be able to pump indefinitely with no risk of overflowing over the top of the glass vessel, and this is because of the top waste line connection. That top waste line should be able to drain liquid at the same rate that the pump delivers the rinse water, but to do this it relies on a clear line with no blockages, running downhill for its full length to the drain or waste container. Often when challenges arise with overflowing, it is due to part of the drain line not running downhill, or running horizontally for too long of a distance. Since it is all gravity-based draining, any place in the line where liquid can settle makes it more difficult for liquid to drain from the vessel. The same thing occurs when growth builds up inside the waste line restricting the flow path.

There is a quick way to test whether overflowing is the source of these challenges:

  1. Remove the top lid from the TitraSip™ vessel (storing the pH electrode in storage solution)
  2. Test the drain valve with the front switch and also drain any liquid that was in the vessel before
  3. Watching the vessel, use the front switch on the TitraSip™ to turn on the Rinse pump. Keep your finger on the switch.
  4. As the vessel fills, it should reach the top connection and start draining down the waste line. The liquid level may go above the connection for a brief second before the full draining starts, but it should not near the top of the glass vessel.
    1. IF it continues filling past the top connection and the liquid reaches the top of the glass vessel, turn off the pump. This would confirm that there is a draining challenge leading to overflows.
    2. IF the draining occurs properly, you should see the liquid level reach a stable level at the top connection, and the pump would be able to be left on indefinitely without overflowing. This would confirm there is NOT a draining challenge and it is likely a leak in the tubing.

How do I prepare my pH electrode for long-term storage or shipping?

To safely store or ship your MANTECH pH electrode, first fill the electrode with fill solution and fill the storage boot ¾ full with fill solution. Then, insert the probe into the boot and screw tight. Place the fill solution plug in its hole and parafilm over the plug, as well as around the body. Then, parafilm over the cable connection spot, making sure not to overlap with the parafilm around the plug.

PeCOD®

Why does KHP measure high on PeCOD®?

The molecule on the surface of the PeCOD® sensor prior to analysis which is a peculiarity of KHP with the PeCOD® COD method. KHP (Potassium Hydrogen Phthalate) has historically been a common reference standard used in a variety of chemistry applications including the traditional dichromate COD test, where it does provide a result close to the theoretical COD result, and for TOC analysis. KHP is not recommended for use in the PeCOD® COD analysis as it over reports compared to the theoretical COD amount. This is predominantly due to some pre-concentration of the molecule on the surface of the PeCOD® sensor prior to analysis which is a peculiarity of KHP with the PeCOD® COD method.

It is important to note that for all COD methods there are specific molecules whereby the individual analytical result is not well aligned to the theoretical value. For instance, organic compounds such as propionic acid, diethylamine or nicotinic acid could not be used as a COD standard for the dichromate COD method due to poor correlation to theoretical results but could be suitable for the PeCOD® COD method. It is therefore important to chose a standard that provides a strong correlation to the theoretical result for the method employed, is a good reflection of the samples to be analyzed, is suitable for general laboratory use and is readily available. For details on preparing sorbitol and glucose-based COD standards for the PeCOD® COD method, read our Technical Bulletin 2017-029: PeCOD® Standard Recipe.

Why does my PeCOD® plot more than one curve during calibration and sample analysis?

For calibration, the PeCOD® Analyzer goes through the following phases:

  1. Normalization Phase – This is where the PeCOD® Analyzer is adjusting the LED lamp output, trying to achieve a baseline of 20µA.
  2. Burn-In Phase (port B) – This is where the pre-mix blank solution is brought into the sensor cell.  It is oxidized to remove contaminants from the cell.
  3. Pre-Burn Phase (port B) – This is where a new aliquot of pre-mix blank solution is brought into the sensor cell.  This phase removed more contaminants from the cell and conditions the cell for the pre-mix blank.
  4. Oxidation (port B) – A new aliquot of pre-mix blank is brought into the sensor cell and is oxidized.  This value is used in the peCOD calculation.  The area under this curve is used to calculate the ‘blank charge’, which corresponds to the C value determined by each calibration. This is the small amount of COD contributed by the electrolyte reagent, which is then subtracted from all future COD analyses to give the final COD result.
  5. Pre-Burn (port A) – A new aliquot of pre-mix calibrant solution is brought into the sensor cell where it is oxidized.  This removes contaminants from the cell and conditions the cell for the pre-mix calibrant.
  6. Oxidation phase (port A) – A new aliquot of pre-mix calibrant if brought into the sensor cell.  This value is used in the peCOD calculation.  The area under the curve is used to calculate the reference charge, which is used with the blank charge to determine the relationship between charge and COD. This corresponds to the M value determined by each calibration.

For sample analysis, the PeCOD® Analyzer goes through the following phases:

  1. Burn-In phase (port B) – The pre-mix blank solution is brought into the sensor cell and is oxidized.  This removed contaminants from the cell.
  2. Pre-Burn phase (port A) – An aliquot of sample is brought into the sensor cell and is oxidized.  This removes contaminants from the cell and conditions the cell for the sample.
  3. Oxidation phase (port A) – A new aliquot of sample is brought into the sensor cell and is oxidized.  This value is used in the peCOD calculation.  The area under the curve is used to calculate the charge generated by sample oxidation.  The blank value will be subtracted from the sample value to determine the sample COD.

Get Smarter Solutions, Today.

Our innovative water and wastewater analyzers have faster, safer processes, with fewer hazardous reagents. Start using MANTECH solutions for more accurate data that drives responsible outcomes.