Why does my turbidity method require a multi-line calibration?

For turbidity, the reasoning is a bit different. The MANTECH turbidity flow cell has specific optical properties which are different than the cuvettes which you would manually insert (for example the sealed standards for the 3-month calibration). The difference in optical properties between the flow cell cylinder and the cuvette can cause the relationship between actual NTU and expected NTU to be slightly non-linear. And like the fluoride, this is really only applicable in the low end of the measuring range, typically <1NTU. To ensure the lowest detection limit possible, we use multi-line to accurately capture how the optical properties of the flow cell are different than the cuvettes.

Why does my fluoride method require a multi-line calibration?

The reasoning for multi-line on Fluoride has to do with the probe response at low levels. We have a stated MDL for the test of 0.01 ppm, and it is known that the slope for fluoride probes (and all ISE probes) deviates from the expected range and approaches zero as the fluoride concentration also approaches zero. We use multi-line to account for this changing slope at the low concentration levels to achieve the lowest detection limit possible. The ability to switch to linear without an effect on results depends on the range you are measuring.