How many PeCOD samples can be analyzed with the reagent starter kit I received?

Home / FAQ / How many PeCOD samples can be analyzed with the reagent starter kit I received?

New PeCOD Analyzer orders typically include a slurry reagent starter kit providing the consumables to run the PeCOD analysis. These kits consist of Electrolyte, Calibrant, and Secondary Standard slurry reagents. Electrolyte is the reagent that is mixed with each sample, therefore it is typically the reagent to be used up first, requiring the user to purchase more from MANTECH. It is useful to know on average how many samples can be run with the electrolyte provided in the standard reagent starter kits, therefore we provide this information below.

You will need to know the color operating range of your PeCOD (Adv. Blue, Green, Yellow, or Red). If you plan to run a single analysis per sample, use the left column for 10mL total volume. If you require duplicate or triplicate analysis per sample, use the right column for 20mL total volume. Depending on how many samples are run per calibration, these numbers may vary.

10mL Total Volume – Single Analysis 20mL Total Volume – Duplicate and Triplicate Analysis
For Advanced Blue range (3:1):

  • 7.5mL sample required
  • 2.5mL electrolyte required
  • 2L electrolyte provided in starter kit
  • 800 samples can be run
For Advanced Blue range (3:1):

  • 15mL sample required
  • 5mL electrolyte required
  • 2L electrolyte provided in starter kit
  • 400 samples can be run
For Green range (1:1):

  • 5mL sample required
  • 5mL electrolyte required
  • 2L electrolyte provided in starter kit
  • 400 samples can be run
For Green range (1:1):

  • 10mL sample required
  • 10mL electrolyte required
  • 2L electrolyte provided in starter kit
  • 200 samples can be run
For Yellow range (1:9):

  • 1mL sample required
  • 9mL electrolyte required
  • 2L provided in starter kit
  • 222 samples can be run
For Yellow range (1:9):

  • 2mL sample required
  • 18mL electrolyte required
  • 2L provided in starter kit
  • 111 samples can be run
For Red range (1:49):

  • 0.2mL sample required
  • 9.8mL electrolyte required
  • 2L provided in starter kit
  • 204 samples can be run
For Red range (1:49):

  • 0.4mL sample required
  • 19.6mL electrolyte required
  • 2L provided in starter kit
  • 102 samples can be run
Number of samples that can be run with 10L bottle of electrolyte:

  • Adv. Blue range, 4000 samples
  • Green range, 2000 samples
  • Yellow range, 1111 samples
  • Red range, 1020 samples
Number of samples that can be run with 10L bottle of electrolyte:

  • Adv. Blue range, 2000 samples
  • Green range, 1000 samples
  • Yellow range, 555 samples
  • Red range, 510 samples
Note: 5x Electrolyte slurry packages are required for making up 10L.

 