What is the flow rate of the IntelliPump?

Home / FAQ / What is the flow rate of the IntelliPump?

The MANTECH IntelliPump is a variable speed intelligent pump that is controlled in speed and direction by the MANTECH Pro software. The pump is set at a specific speed (1-100) and either 1/4″ or 1/8″ diameter tubing is used to achieve desired flow rates for various applications. The pump can then be calibrated by measuring liquid dispensed over a set period of time, and entering the value into the Hardware Configuration window of the software.

Pumps are assigned a default estimated flow rate based on the speed setting and tubing size, and this flow rate is used unless the end user decides to run a pump calibration. The estimated flow rates for various speed and tubing settings are provided below:

Tubing Diameter (inches) Speed Setting (%) Flow Rate (mL/min)
1/8” 6 (min) 1.9
10 17.88
20 56.22
30 92.27
40 126.02
50 157.47
60 186.63
70 213.49
80 239.05
90 260.32
98 (max) 276.48
1/4” 3 (min) 2.79
10 61.82
20 146.15
30 230.48
40 314.81
50 399.14
60 483.47
70 567.80
80 652.13
90 736.46
99 (max) 812.36