What is the minimum total liquid volume that can be measured for pH, conductivity and alkalinity with the MT Systems?

Home / FAQ / What is the minimum total liquid volume that can be measured for pH, conductivity and alkalinity with the MT Systems?

When using 50mL sample tubes, the minimum total volume can be as small as 6 ml using the TitraPro4 pH electrode and the MANTECH 5mm conductivity probe. This is due to the lower immersion depth of these electrodes and precise autosampler coordinate specifications. When using 125mL sample cups, the minimum total volume can be as small as 15mL.

 