T10 Technical Specifications

This document outlines the technical specification for the T10 turbidity meter. Please note that these specifications apply to manual turbidity analysis with the T10 meter only.

 

For details about automated turbidity analysis with the T10 and MT Environmental Titrators, please contact MANTECH at 519-763-4245 or at info@mantech-inc.com.

 

Measurement method Ratio determination using a primary nephelometric light scatter signal (90°) to the transmitted light scatter signal.
Reading units NTU/ EBC
Lamp Source Tungsten Lamp (White Light); LED Lamp (IR Light)
Method Conformity EPA 180.1, ASTM D1889, SM 2130B (White Light);
ISO 7027 and EN27027 (IR Light)
Detector Silicon Photocell
Measuring Range 0 to 1000 NTU / 0 to 250 EBC
Resolution 0.01 on lowest range
Accuracy ± 2% of reading: 0 to 500 NTU / ± 3% 500 to 1000 NTU
Repeatability ± 1% of reading or 0.01 NTU
Automatic Reading With user-defined intervals 0 to 250 seconds
Maximum uncertainty ± 2% of full scale
Display LCD 2 lines / 16 characters
Response Time Programmable 6 to 41 sec
Data Logger Up to 1000 data
Auto ShutOff Programmable from 1 to 60 min
Fast Cal function Quick Calibration for single point
Software Functions Signal averaging, “Fast Settling”, results freezing, analyst and sample identification, calibration status, verification and calibration reminder, calibration history, password
Sample Required 15 ml vial with lid
Sample vials Round borosilicate glass vial with screw and caps (Φ = 24,5 mm)
Power Supply 4 AA batteries or USB power supply cable
Indicator Low battery indicator / battery exchange
Serial Output USB
Storage Conditions 0 to 40°C (instrument only)
Dimensions 114 x 198 x 83 mm

For the PDF, please click here:  MAN-MD-H-0507-03 MANTECH Turbidity T10 Meter Technical Specifications